specifications 8753ET /ES Network Analyzers HP 8753ET and 8753ES Network Analyzers 30 kHz to 3 or 6 GHz This document describes the performance and features of the following products: HP 8753ES S-parameter vector network analyzer HP 8753ES Option 011 vector network analyzer without test set HP 8753ET T ransmission/reflection vector network analyzer For more information about these analyzers, please see the following documents: HP literature number HP 8753ET and 8753ES Network Analyzers Brochure 5968-5159E HP 8753ET and 8753ES Network Analyzers Configuration Guide 5968-5158E
2 T able of contents Definitions and test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 System performance summaries . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Specifications and characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Measurement throughput summary . . . . . . . . . . . . . . . . . . . . . . . . .21 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 T est set specifications for Option 011 . . . . . . . . . . . . . . . . . . . . . . . .22 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
3 Definitions and test conditions Specifications describe the instrument's warranted performance after a half-hour warm-up and over the temperature range of 25 ˚C ± 5 ˚C, unless otherwise stated. Specifications for frequencies above 3 GHz do not apply to instruments with Option 075 (75-ohm impedance). Supplemental characteristics are typical but non-war- ranted performance parameters. These are denoted as “typical,” “nominal,” or “approximate.” The measurement uncertainty curves and measure- ment port characteristics given for HP 8753ES systems also apply to the HP 8753ES with Options 006 and 011 and the HP 85047A test set (50-ohm), or the HP 8753ES Option 011 with an HP 85046B test set (75-ohm). Dynamic range System dynamic range is calculated as the difference between the receiver noise floor and the lesser of either the source maximum output or the receiver maximum input level. System dynamic range applies to transmission measurements only , since reflection mea- surements are limited by directivity . Noise floor is specified as the mean of the noise trace over frequency . Noise floor is measured with the test ports terminated in loads, full two-port error correc- tion for the HP 8753ES and enhanced-response error correction for the HP 8753ET (with 16 averages used during isolation), 10 Hz IF bandwidth (BW), maximum test port power , and no averaging during the measure- ment. Measurement uncertainty Measurement uncertainty curves utilize a Root Sum Square (RSS) model for the contribution of random errors such as noise, typical connector repeatabilities, and test set switching; this is combined with a worst- case model for the contributions of dynamic accuracy and residual systematic errors. Curves show the worst-case magnitude and phase uncertainty for reflection and transmission measure- ments, after a full two-port error correction for the HP 8753ES and enhanced-response error correction for the HP 8753ET (including isolation with an averag- ing factor of 16) using the specified cal kit, with 10 Hz IF bandwidth (BW) and no averaging. Measurement port characteristics Characteristics show the residual system uncertainties for uncorrected performance and after accuracy enhancement using full two-port error correction for the HP 8753ES and enhanced-response error correc- tion for the HP 8753ET . These characteristics apply for an environmental temperature of 25 ± 5 ˚C, with less than 1 ˚C deviation from the calibration tempera- ture. HP 8753ET Option 004 may degrade transmission source match as much as 2 dB, resulting in up to 0.05 dB additional uncertainty in transmission tracking. Corrected performance indicates residual error after calibration. It is determined by the quality of calibra- tion standards, system repeatability , stability , and noise. Uncorrected performance indicates intrinsic errors without calibration correction applied. This is related to the ultimate stability of a calibration.
4 System performance summaries HP 8753ES (50-ohm systems) 7-mm test ports Configuration Network analyzer HP 8753ES Standard and Option 006 Calibration kit HP 85031B T est-port cables HP 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Reflection uncertainty Magnitude T ransmission uncertainty Magnitude Measurement port characteristics Frequency range Corrected 30 kHz-300 kHz 2 300 kHz-1.3 GHz 1.3 GHz-3 GHz 3 GHz- 6 GHz Directivity 55 dB 55 dB 51 dB 46 dB Source match 55 dB 51 dB 49 dB 43 dB Load match 55 dB 55 dB 51 dB 46 dB Reflection ±(0.001 dB ±(0.001 dB ±(0.005 dB ±(0.020 dB tracking 0.02 dB/˚C) 0.01 dB/˚C) 0.02 dB/˚C) 0.03 dB/˚C) T ransmission ±(0.008 dB ±(0.006 dB ±(0.009 dB ±(0.021 dB tracking 0.02 dB/˚C) 0.01 dB/˚C) 0.02 dB/˚C) 0.03 dB/˚C) Uncorrected Directivity 20 dB* 35 dB 30 dB 25 dB Source match 18 dB** 16 dB 16 dB 14 dB Load match 18 dB** 18 dB 16 dB 14 dB Reflection ±2.5 dB ±1 dB ±1 dB ±1.5 dB tracking T ransmission ±2.5 dB ±1 dB ±1 dB ±1.5 dB tracking Crosstalk 90 dB*** 100 dB 100 dB 90 dB *15 dB, 30 kHz to 50 kHz **10 dB, 30 kHz to 50 kHz ***60dB, 30 kHz to 50 kHz Phase Phase 1. T ypical below 300 kHz. 2. T ypical performance.
5 System performance summaries HP 8753ES (50-ohm systems) T ype-N test ports Configuration Network analyzer HP 8753ES Standard and Option 006 Calibration kit HP 85032B T est-port cables HP 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Reflection uncertainty Magnitude T ransmission uncertainty Magnitude Measurement port characteristics Frequency range Corrected 30 kHz-300 kHz 2 300 kHz-1.3 GHz 1.3 GHz-3 GHz 3 GHz-6 GHz Directivity 50 dB 50 dB 47 dB 40 dB Source match 49 dB 42 dB 36 dB 31 dB Load match 50 dB 50 dB 47 dB 40 dB Reflection ±(0.005 dB ±(0.009 dB ±(0.019 dB ±(0.070 dB tracking 0.02 dB/˚C) 0.01 dB/˚C) 0.02 dB/˚C) 0.03 dB/˚C) T ransmission ±(0.014 dB ±(0.013 dB ±(0.026 dB ±(0.065 dB tracking 0.02 dB/˚C) 0.01 dB/˚C) 0.02 dB/˚C) 0.03 dB/˚C) Phase Phase 1. T ypical below 300 kHz. 2. T ypical performance.
6 System performance summaries HP 8753ES (50-ohm systems) 3.5-mm test ports Configuration Network analyzer HP 8753ES Standard and Option 006 Calibration kit HP 85033D T est-port cables HP 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Reflection uncertainty Magnitude T ransmission uncertainty Magnitude Measurement port characteristics Frequency range Corrected 30 kHz-300 kHz 2 300 kHz-1.3 GHz 1.3 GHz-3 GHz 3 GHz-6 GHz Directivity 49 dB 46 dB 44 dB 38 dB Source match 49 dB 44 dB 41 dB 37 dB Load match 49 dB 46 dB 44 dB 38 dB Reflection ±(0.010 dB ±(0.005 dB ±(0.007 dB ±(0.009 dB tracking 0.02/˚C) 0.01/˚C) 0.02/˚C) 0.03/˚C) T ransmission ±(0.016 dB ±(0.014 dB ±(0.022 dB ±(0.048 dB tracking 0.02/˚C) 0.01/˚C) 0.02/˚C) 0.03/˚C) Phase Phase 1. T ypical below 300 kHz. 2. T ypical performance.
7 System performance summaries HP 8753ES (75-ohm systems) T ype-N test ports Configuration Network analyzer HP 8753ES Option 075 Calibration kit HP 85036B T est-port cables HP 11857B System dynamic range 30 kHz to 50 kHz 68 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 96 dB 16 MHz to 3 GHz 106 dB Reflection uncertainty Magnitude T ransmission uncertainty Magnitude Measurement port characteristics Frequency range Corrected 30 kHz - 300 kHz 2 300 kHz - 1.3 GHz 1.3 GHz - 3 GHz Directivity 48 dB 48 dB 43 dB Source match 47 dB 41 dB 35 dB Load match 48 dB 48 dB 43 dB Reflection ±(0.004 dB ±(0.010 dB ±(0.019 dB tracking 0.02 dB/˚C) 0.01 dB/˚C) 0.02 dB/˚C) T ransmission ±0.018 dB ±(0.016 dB ±(0.033 dB tracking 0.02 dB/˚C) 0.01 dB/˚C) 0.02 dB/˚C) Uncorrected 2 30 kHz - 300 kHz 300 kHz - 1.3 GHz 1.3 GHz - 3 GHz Directivity 20 dB* 35 dB 30 dB Source match 16 dB** 16 dB 16 dB Load match 15 dB** 18 dB 16 dB Reflection ±2.5 dB ±1 dB ±1 dB tracking T ransmission ±2.5 dB ±1 dB ±1 dB tracking Crosstalk 90 dB*** 100 dB 100 dB * 15 dB, 30 to 50 kHz **10 dB, 30 to 50 kHz *** 60 dB, 30 to 50 kHz Phase Phase 1. T ypical below 300 kHz. 2. T ypical performance.
8 System performance summaries HP 8753ES (75-ohm systems) T ype-F test ports Configuration Network analyzer HP 8753E Option 075 Calibration kit HP 85039B T est-port cables HP 11857B System dynamic range 30 kHz to 50 kHz 68 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 96 dB 16 MHz to 3 GHz 106 dB Reflection uncertainty Magnitude T ransmission uncertainty Magnitude Measurement port characteristics Data is shown for a T ype-F female reflection port and a T ype-F male transmission port. Frequency range Corrected 2 30 kHz-300 kHz 300 kHz-1.3 GHz 1.3 GHz-3 GHz Directivity 45 dB 45 dB 40 dB Source match 40 dB 40 dB 30 dB Load match 45 dB 45 dB 40 dB Reflection ±(0.060 dB ±(0.060 dB ±(0.024 dB tracking 0.02 dB/˚C) 0.01 dB/˚C) 0.02 dB/˚C) T ransmission ±(0.033 dB ±(0.019 dB ±(0.057 dB tracking 0.02 dB/˚C) 0.01 dB/˚C) 0.02 dB/˚C) Phase Phase 1. T ypical below 300 kHz. 2. T ypical performance.
9 System performance summaries HP 8753ET (50-ohm systems) T ype-N test ports Configuration Network analyzer HP 8753ET Standard, Option 006 or Option 004 Calibration kit HP 85032B T est port cable HP part number 8120-4781 System dynamic range 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB 3 GHz to 6 GHz 105 dB Measurement port characteristics Frequency range Uncorrected 300 kHz-1.3 GHz 1.3 GHz-3.0 GHz 3.0 GHz-6.0 GHz Directivity 30 dB 24 dB 19 dB Source match 4 25 dB 20 dB 14 dB Load match 24 dB 19 dB 16 dB Reflection tracking ±1.0 dB ±1.0 dB ±2.0 dB T ransmission tracking ±1.5 dB ±1.5 dB ±2.5 dB Crosstalk 100 dB 100 dB 90 dB Frequency range (continued) Corrected 300 kHz-1.3 GHz 1.3 GHz-3.0 GHz 3.0 GHz-6.0 GHz Reflection measurements 1 Directivity 50 dB 47 dB 40 dB Source match 42 dB 36 dB 31 dB Load match One-port cal 24 dB 19 dB 16 dB Enhanced reflection cal 24 dB 19 dB 16 dB Reflection tracking ±(0.009 dB ±(0.019 dB ±(0.07 dB 0.01 dB/˚C) 0.02 dB/˚C) 0.03 dB/˚C) T ransmission measurements 2 Source match Enhanced-response cal 42 dB 36 dB 31 dB Response-only cal 3 25 dB 20 dB 14 dB T ransmission tracking Enhanced-response cal ±(0.006 dB ±(0.018 dB ±(0.054 dB 0.01 dB/˚C) 0.02 dB/˚C) 0.03 dB/˚C) Response-only cal ±(0.033 dB ±(0.1 dB ±(0.27 dB 0.01 dB/˚C) 0.02 dB/˚C) 0.03 dB/˚C) 1. One-port or enhanced-response calibration. 2. Enhanced-response or response-only calibration. 3. 2 dB less with Option 004. 4. Option 004 may degrade uncorrected source match as much as 2 dB. T ransmission uncertainty: (enhanced-response calibration) T ransmission uncertainty: (response calibration) Reflection uncertainty (one-port calibration)
10 HP 8753ES specifications T est-port output characteristics 7 Frequency characteristics Range 30 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability 9 ±7.5 ppm (0˚ to 55 ˚C) ±3 ppm/year Accuracy ±10 ppm at 25 ˚C ± 5 ˚C Option 1D5 ±0.05 ppm (0˚ to 55 ˚C), ±0.5 ppm/year Power range 2 -85 to 10 dBm Option 075/014 -85 to 8 dBm Power sweep range 25 dB (typically 31 dB) Resolution 0.01 dB Level accuracy 1,2,5 ± 1.0 dB Level linearity 1,2,5 -15 dBm to 5 dBm ±0.2 dB 5 dBm to 10 dBm 6 ±0.5 dB Impedance 9 50 Ω (nominal) 30 kHz to 3 GHz >16 dB RL (<1.38 SWR) 3 GHz to 6 GHz >14 dB RL (<1.50 SWR) Option 075 75 Ω ( nominal ) 30 kHz to 3 GHz >16 dB RL (<1.38 SWR) Spectral purity 2nd harmonic 3 <-40 dBc at 0 dBm 9 <-50 dBc at -10 dBm 9 Option 002 <-25 dBc at 10 dBm 6 3rd harmonic 4 <-40 dBc at 0 dBm 9 <-50 dBc at -10 dBm 9 Option 002 <-25 dBc at 10 dBm 6 Nonharmonic spurious 9 Mixer related <-30 dBc at 10 dBm 6 <-55 dBc at -10 dBm T est port input characteristics Frequency range 30 kHz to 3 GHz Option 006 30 kHz to 6 GHz A verage noise level 2,8 <3 GHz <-82 dBm (3 kHz BW) <3 GHz <-102 dBm (10 Hz BW) <3 GHz <-110 dBm (10 Hz BW) typical 3 to 6 GHz <-77 dBm (3 kHz BW) 3 to 6 GHz <-97 dBm (10 Hz BW) 3 to 6 GHz <-105 dBm (10 Hz BW) typical Maximum input level 10 dBm Damage level 26 dBm or 35 VDC Impedance 50 Ω (nominal) Option 075 75 Ω ( nominal ) Frequency response 2,5 300 kHz to 3 GHz ±1.0 dB 3 GHz to 6 GHz ±2.0 dB Internally generated harmonics (Option 002) 2nd harmonic 3 <-15 dBc at 8 dBm <-30 dBc at 0 dBm 9 <-45 dBc at -15 dBm 9 3rd harmonic 4 <-30 dBc at 8 dBm <-50 dBc at 0 dBm 9 <-50 dBc at -15 dBm 9 Harmonic measurement accuracy 10 300 kHz to 3 GHz ±1.5 dB 3 GHz to 6 GHz ±3 dB (Option 006) Harmonic measurement dynamic range -40 dBc (output = -10 dBm, input = <-15 dBm) 9 Frequency offset mode 11 Frequency range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz R-channel input requirements Power level 300 kHz to 3 GHz 0 to -35 dBm 3 GHz to 6 GHz 0 to -30 dBm LO spectral purity (typical) Maximum <-25 dBc spurious input Residual FM <20 kHz LO frequency accuracy 9 ± 1 MHz at nominal frequency 1. Relative to 0 dBm output power . 2. T ypical below 300 kHz. 3. 16 MHz to 3 GHz. 4. 16 MHz to 2 GHz. 5. T ypical from 2 to 3 GHz for instruments with Option 075. 6. 8 dBm maximum with Option 075, or Option 014. 7. T est performed on port 1 only . 8. Instruments with Option 075 are degraded 2 dB. 9. T ypical performance. 10. 25˚ ± 5˚ C. 11. The HP 8753ES source characteristics and measurement accuracy in this mode are dependent on the stability of the external LO source. The RF source tracks the LO to maintain a stable IF signal at the R-channel receiver input. Degradation in accuracy is negligible when using an HP 8642A/B, 8656B, or E4432B RF signal generator as the LO source.
11 T est port input characteristics (continued) External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements (typical) Power level 0 to -25 dBm Spectral purity Maximum <-30 dBc spurious input Residual FM <20 kHz T ypical settling 500 ms (automatic) time 50 ms (manual) Frequency readout 0.1% (automatic) accuracy Input frequency accuracy requirement 2 Manual -0.5 to 5 MHz Accuracy (see magnitude and phase characteristics) Magnitude characteristics Dynamic accuracy (10 Hz IF BW) Display resolution 0.001 dB/division Marker resolution 3 0.001 dB T race noise ( 5 dBm at test-port, ratio measurement, 3 kHz BW) 300 kHz to 3 GHz < 0.006 dB rms 3 GHz to 6 GHz < 0.010 dB rms Reference level Range ± 500 dB Resolution 0.001 dB Stability 2 30 kHz to 3 GHz 0.02 dB/ ˚C 3 GHz to 6 GHz 0.04 dB/ ˚C Phase characteristics Dynamic accuracy (10 Hz IF BW) Range ±180˚ Display resolution 0.01˚/ division Marker resolution 3 0.01˚ T race noise ( 5 dBm at test-port, ratio measurement, 3 kHz BW) 300 kHz to 3 GHz < 0.038˚ rms 3 GHz to 6 GHz < 0.070˚ rms Reference level Range -180˚ to 180˚ Resolution 0.01˚ Stability 30 kHz to 3 GHz 0.05˚/ ˚C 3 GHz to 6 GHz 0.20˚/ ˚C Polar characteristics Range 10 x 10 -12 to 1000 units full scale Reference ±500 units 1. See the HP 8753ES descriptions and options for a functional description. Measurement accuracy is dependent on the stability of the input signal. 2. T ypical performance. 3. Marker resolution for magnitude; phase and delay is dependent upon measured value. Resolution is limited to five digits. (300 kHz to 3GHz) (3 to 6GHz) (300 kHz to 3GHz) (3 to 6GHz)
12 HP 8753ES Option 011 specifications T est port output characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability typically ±7.5 ppm 0˚ to 55 ˚C typically ±3 ppm/year Option 1D5 typically ± 0.05 ppm 0˚ to 55 ˚C typically ±0.5 ppm/year Accuracy ±10 ppm at 25 ˚C ± 5 ˚C Power range -5 to 20 dBm Option 006 -5 to 18 dBm Resolution 0.01 dB Level accuracy 1,2 ± 1.0 dB Level linearity 1,2,5 ±0.25 dB, -5 to 15 dBm ±0.5 dB, 15 to 20 dBm Impedance 50 Ω nominal 300 kHz to 3 GHz 2 >16 dB RL (<1.38 SWR) 300 kHz to 6 GHz >14 dB RL (<1.50 SWR) Spectral purity 2nd harmonic 3 <-40 dBc at 10 dBm 6 <-50 dBc at 0 dBm 6 Option 002 <-25 dBc at max power 3rd harmonic 4 <-40 dBc at 10 dBm 6 <-50 dBc at 0 dBm 6 Option 002 <-25 dBc at max power Nonharmonic spurious 6 Mixer related <-30 dBc at max power <-55 dBc at -10 dBm T est port input characteristics Option 011 Frequency range 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz A verage noise level 2 50 kHz to 3 GHz <-90 dBm (3 kHz BW) 50 kHz to 3 GHz <-110 dBm (10 Hz BW) 50 kHz to 3 GHz <-120 dBm (10 Hz BW) typical 3 to 6 GHz <-85 dBm (3 kHz BW) 3 to 6 GHz <-105 dBm (10 Hz BW) 3 to 6 GHz <-115 dBm (10 Hz BW) typical Maximum input level 0 dBm Damage level 20 dBm or 25 VDC Impedance: 50 ohms nominal 300 kHz to 2 MHz ≥20 dB RL 2 MHz to 1.3 GHz ≥24 dB RL 1.3 GHz to 3 GHz ≥19 dB RL Option 006 3 GHz to 6 GHz ≥15 dB RL 6 Frequency response 300 kHz to 3 GHz ±1.0 dB, 3 GHz to 6 GHz ±2.0 dB Harmonics (Option 002) 2nd harmonic 3 <-15 dBc at 0 dBm <-30 dBc at -10 dBm 6 <-45 dBc at -30 dBm 6 3rd harmonic 4 <-30 dBc at 0 dBm <-50 dBc at -10 dBm 6 <-50 dBc at -30 dBm 6 Harmonic measurement accuracy 16 MHz to 3 GHz ±1.5 dB 3 GHz to 6 GHz Option 006 ±3 dB Harmonic measurement dynamic range -40 dBc 6 (output = -10 dBm, input <-15 dBm) 1. 10 dBm output power for the HP 8753ES Option 011. 2. T ypical below 300 kHz. 3. 16 MHz to 3 GHz. 4. 16 MHz to 2 GHz. 5. For HP 8753ES Option 011 and Option 006, linearity is specified for the ranges of -5 to 13 dBm and 13 to 18 dBm. 6. T ypical performance.
13 HP 8753ES Option 011 specifications T est port input characteristics (continued) Frequency offset mode 3 Frequency range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz R-channel input requirements Power level 300 kHz to 3 GHz 0 to -35 dBm 3 GHz to 6 GHz 0 to -30 dBm LO spectral purity Maximum spurious <-25 dBc input Residual FM <20 kHz LO frequency accuracy ± 1 MHz at nominal frequency External source mode 4 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements 1 Power level 0 to -25 dBm Spectral purity Maximum spurious <-30 dBc input Residual FM <20 kHz T ypical settling time 1 500 ms (automatic) 50 ms (manual) Input frequency accuracy requirement 1 Manual: -0.5 to 5 MHz Display resolution 0.001 dB/division Marker resolution 5 0.001 dB T race noise ( 5 dBm at test-port, ratio measurement, 3 kHz BW) Magnitude 300 kHz to 3 GHz < 0.006 dB rms 3 GHz to 6 GHz < 0.010 dB rms Phase 300 kHz to 3 GHz < 0.038 ˚ rms 3 GHz to 6 GHz < 0.070 ˚ rms Reference level Range ± 500 dB Resolution 0.001 dB Stability 2 30 kHz to 3 GHz 0.02 dB/ ˚C 3 GHz to 6 GHz 0.04 dB/ ˚C Range ±180˚ Display resolution 0.01˚/ division Marker resolution 5 0.01˚ T race noise 2 30 kHz to 3 GHz < 0.038˚ rms 3 GHz to 6 GHz < 0.070˚ rms ( 5 dBm at test port, ratio measurement, 3 kHz BW) Reference level Range -180˚ to 180˚ Resolution 0.01˚ Stability 30 kHz to 3 GHz 0.05˚/ ˚C 3 GHz to 6 GHz 0.20˚/ ˚C Polar characteristics Range 10 x 10 -12 to 1000 units full scale Reference ± 500 units 1. T ypical performance. 2. T ypical below 300 kHz. 3. The HP 8753ES source characteristics and measurement accuracy in this mode are dependent on the stability of the external LO source. The RF source tracks the LO to maintain a stable IF signal at the R-channel receiver input. Degradation in accuracy is negligible when using an HP 8642A/B, 8656B, or E4432B RF signal generator as the LO source. 4. See the HP 8753ES descriptions and options for a functional description. Measurement accuracy is dependent on the stability of the input signal. 5. Marker resolution for magnitude, phase and delay is dependent upon measured value. Resolution is limited to five digits.
14 HP 8753ET specifications T est-port output characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz Resolution 1 Hz Stability 4 ±7.5 ppm (0˚ to 55 ˚C) ±3 ppm/year Option 1D5 ± 0.05 ppm (0˚ to 55 ˚C) ±0.5 ppm/year Accuracy ±10 ppm at 25 ˚C ±5 ˚C Power range: -20 to 5 dBm -85 to 10 dBm (with Option 004) Resolution 0.01 dB Level accuracy 1 ± 1.0 dB Level linearity 1 Standard -20 to -15 dBm ±0.5 dB -15 to 0 dBm ±0.2 dB 0 to 5 dBm ±0.5 dB Option 004 -15 to 5 dBm ±0.2 dB 5 to 10 dBm ±0.5 dB Impedance 50 Ω (nominal) 300 kHz to 3 GHz >18 dB RL (< 1.28 SWR) 3 GHz to 6 GHz > 14 dB RL (<1.50 SWR) Spectral purity 2nd harmonic 2 <-40 dBc at 0 dBm 4 <-50 dBc at -10 dBm 4 Option 002 <-25 dBc at maximum output power 3rd harmonic 3 <-40 dBc at 0 dBm <-50 dBc at -10 dBm Option 002 <-25 dBc at maximum output power Nonharmonic spurious Mixer related <-30 dBc at 10 dBm <-55 dBc at -10 dBm T est port input characteristics Frequency range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz A verage noise level 300 kHz to 3 GHz <-90 dBm (3 kHz BW) 300 kHz to 3 GHz <-110 dBm (10 Hz BW) 300 kHz to 3 GHz <-120 dBm (10 Hz BW) typical 3 GHz to 6 GHz <-85 dBm (3 kHz BW) 3 GHz to 6 GHz <-105 dBm (10 Hz BW) 3 GHz to 6 GHz <-114 dBm (10 Hz BW) typical Maximum input level 10 dBm reflection port 0 dBm transmission port Damage level > 26 dBm or 35 VDC Impedance 50 Ω (nominal) Frequency response 5 300 kHz to 3 GHz ±1.0 dB 3 GHz to 6 GHz ±2.0 dB Harmonics (Option 002) 2nd harmonic 2 <-15 dBc at 8 dBm <-30 dBc at 0 dBm 4 <-45 dBc at -15 dBm 4 3rd harmonic 3 <-30 dBc at 8 dBm <-50 dBc at 0 dBm 4 <-50 dBc at -15 dBm 4 Harmonic measurement accuracy 5 300 kHz to 3 GHz ±1.5 dB 3 GHz to 6 GHz ±3 dB (Option 006) Harmonic measurement dynamic range 4 -40 dBc (output = -10 dBm, input = <-15 dBm) Frequency offset mode 6 Frequency range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz R-channel input requirements Power level 300 kHz to 3 GHz 0 to -35 dBm 3 GHz to 6 GHz 0 to -30 dBm LO Spectral purity Maximum spurious <-25 dBc input Residual FM <20 kHz LO frequency accuracy 6 ± 1 MHz at nominal frequency 1. Relative to -5 dBm output power for the HP 8753ET ; -10 dBm output power for HP 8753ET with Option 004. 2. 16 MHz to 3 GHz. 3. 16 MHz to 2 GHz. 4. T ypical performance. 5. 25˚ C ±5˚ C. 6. The HP 8753ET source characteristics and measurement accuracy in this mode are dependent on the stability of the external LO source. The RF source tracks the LO to maintain a stable IF signal at the R-channel receiver input. Degradation in accuracy is negligible when using an HP 8642A/B, 8656B, or E4432B RF signal generator as the LO source.
15 HP 8753ET specifications T est-port input characteristics (continued) External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements 2 Power level 0 to -25 dBm Spectral purity Maximum spurious <-30 dBc input Residual FM <20 kHz T ypical settling time 500 ms (automatic) 50 ms (manual) Frequency readout 0.1% (automatic) accuracy 2 Input frequency accuracy requirement 2 Manual: -0.5 to 5 MHz Accuracy (See magnitude and phase characteristics) Magnitude characteristics Dynamic accuracy (10 Hz IF BW) Display resolution 0.001 dB/division Marker resolution 3 0.001 dB T race noise (0 dBm at transmission port or 5 dBm at reflection port, ratio measurement, 3 kHz BW) 300 kHz to 3 GHz < 0.006 dB rms 3 GHz to 6 GHz < 0.010 dB rms Reference level Range ± 500 dB Resolution 0.001 dB Stability 2 300 kHz to 3 GHz 0.02 dB/ 1 ˚C 3 GHz to 6 GHz 0.04 dB/ 1 ˚C Phase characteristics Dynamic accuracy (10 Hz IF BW) Range ±180˚ Display resolution 0.01˚/ division Marker resolution 3 0.01˚ T race noise (0 dBm at transmission port or 5 dBm at reflection port, ratio measurement, 3 kHz BW) 300 kHz to 3 GHz < 0.038˚ rms 3 GHz to 6 GHz < 0.070˚ rms Reference level Range -180˚ to 180˚ Resolution 0.01˚ Stability 30 kHz to 3 GHz 0.05˚/ ˚C 3 GHz to 6 GHz 0.20˚/ ˚C Polar characteristics Range 10 x 10 -12 to 1000 units full scale Reference ± 500 units 1. See the HP 8753ET/ES descriptions and options for a functional description. Measurement accuracy is dependent on the stability of the input signal. 2. T ypical performance. 3. Marker resolution for magnitude, phase and delay is dependent upon measured value. Resolution is limited to five digits. (300 kHz to 3 GHz) (3 to 6 GHz) (3 to 6 GHz) (300 kHz to 3 GHz)
16 HP 8753ET/ES supplemental characteristics Measurement Number of display channels Four display channels available. Number of measurement channels T wo primary and two auxiliary measurement channels available. Measurement parameters HP 8753ET : Reflection, transmission, A, B, R, A/R, B/R, A/B. Conversion to impedance or admittance. HP 8753ES: S11, S21, S12, S22, A, B, R, A/R, B/R, A/B. Conversion to impedance or admittance. Formats Cartesian: log/linear magnitude, phase, group delay , SWR, real and imaginary . Smith chart: with log/linear amplitude and phase, R jX, G jB, or real/imaginary markers. Polar: with linear/log amplitude, phase, or real and imaginary markers. Data markers Each display channel has five independent markers that can be displayed simultaneously . T wenty indepen- dent markers can be displayed in 4-channel display mode when markers are uncoupled. Marker functions Markers can be used in various functions: Marker search (Mkr to max, Mkr to min, Mkr to target), Mkr bandwidth with user -defined target values, Mkr to start, Mkr to stop, Mkr to center , Mkr to span, Mkr to reference, Mkr to delay , and trace statistics (average value, standard deviation, and peak-to-peak deviation of the data trace between two markers). The tracking function enables continuous update of marker search values on each sweep. Group delay characteristics Aperture: selectable Maximum aperture: 20% of frequency span Minimum aperture: (freq. span) / (number of points 1) Range The maximum delay is limited to measuring no more than 180˚ of phase change within the minimum aper- ture. Range = 1 / (2 x minimum aperture) Accuracy The following graph shows group-delay accuracy at 1.3 GHz with type-N full two-port calibration and 10-Hz IF bandwidth. Insertion loss is assumed to be < 2 dB and electrical length to be ten meters. Source control Sweep limits Set start/stop or center/span of the stimulus parameter (frequency , power , time) directly through the source control keys and the control knob, the step keys or the data entry keyboard. Sweep type Set a linear or logarithmic sweep, an arbitrarily defined frequency list, a power sweep or a CW (single frequen- cy) type of sweep. Measured number of points per sweep Linear frequency: choose 3, 11, 26, 51, 101, 201, 401, 801, or 1601 points. Fast swept list Define up to 30 different sub-sweep frequency ranges in any combination of CW , center/span, or start-stop sweep modes. Set test-port power levels and IF band- width independently for each segment. Sweep modes Set a coupled channel sweep (same stimulus condi- tions on both channels) or an uncoupled channel sweep. Chop/alternate Select whether to alternately or simultaneously (chop) measure channels when in dual-channel mode. Chop mode is faster , while alternate mode optimizes dynam- ic range. Sweep time Set sweep time in seconds, minutes or hours. Automatic sweep time Select auto sweep time by entering zero seconds sweep time. The analyzer will sweep at the minimum sweep time for any subsequently selected stimulus conditions. Auto sweep time is the default condition.
17 HP 8753ET/ES supplemental characteristics (continued) Sweep trigger Set to either continuous, single, group sweep, or exter- nal trigger . Set external trigger to take a complete sweep or to measure individual points in a frequency , power or list sweep. Power Set source power from -20 to 5 dBm for the HP 8753ET 1 or from -85 to 10 dBm for the HP 8753ES 2 . Power slope can be set in dBm/GHz. Power meter calibration Select continuous leveling or use a correction table to modify source power . The correction table is created with an initial single sweep. Make single or multiple power meter readings at each frequency . Data accuracy enhancement Measurement calibration Measurement calibration significantly reduces mea- surement uncertainty due to errors caused by system directivity , source and load match, tracking, and crosstalk. Full two-port calibration removes all the sys- tematic errors to obtain the most accurate measure- ments. Calibration types available • Frequency response Simultaneous magnitude and phase correction of fre- quency response errors for either reflection or trans- mission measurements. • Response and isolation Compensates for frequency response and directivity (reflection) or frequency response and crosstalk errors. • Enhanced response calibration Corrects for frequency response and source match for transmission measurements, and provides one-port calibration for reflection measurements. • One-port calibration Uses test set port 1 or port 2 to correct for directivity , frequency response and source match errors. • T wo-port calibration Compensates for directivity , source match, reflection frequency response, load match, transmission frequen- cy response and crosstalk for an S-parameter test set. Crosstalk calibration can be omitted. A vailable on HP 8753ES analyzers. • TRL*/LRM* calibration Compensates for directivity , reflection and transmis- sion frequency response, and crosstalk in both the for- ward and reverse directions. Especially suitable for calibrating non-coaxial environments, such as in test fixtures. TRL*/LRM* is a special implementation of TRL/LRM calibration, modified for the three-sampler receiver in the HP 8753ES. A vailable on HP 8753ES analyzers. Interpolated error correction W ith any type of accuracy enhancement applied, inter- polated mode recalculates the error coefficients when the test frequencies are changed. The number of points can be increased or decreased and the start/stop frequencies can be changed, but the result- ing frequency span must be a subset of the original cal- ibration frequency span. System performance is not specified for measurements with interpolated error correction applied. V elocity factor Enters the velocity factor to calculate equivalent elec- trical length. Reference plane extension Redefine the plane-of-measurement reference to other than port 1 or port 2 of the HP 8753ET and 8753ES. Select default calibration kit Select from a list of standard calibration kits or choose a user -defined kit. Data averaging IF bandwidth: The IF bandwidth is selectable from 6 kHz to 10 Hz W eighted sweep-to-sweep averaging: A verages vector data on each successive sweep. T race smoothing Computes the moving average of adjacent data points. Smoothing aperture defines the trace width (number of points) to be averaged, and ranges from 0.25% to 20% of the trace width. Display control Display formats Single-channel, dual-channel overlay (both traces on one graticule), dual-channel split (each trace on sepa- rate graticules), three-channel split (each trace on separate graticules), three-channel overlay (three traces on one graticule), quad-channel overlay (four traces on one graticule), quad-channel split (each trace on separate graticules). 1. -85 to 5 dBm with Option 004. 2. 8 dBm maximum with Option 075 or 014.
18 HP 8753ET/ES supplemental characteristics (continued) T race functions Display data Display current measurement data, memory data, or current measurement with measurement and memory data simultaneously . T race math V ector division or subtraction of current linear mea- surement values and memory data. Display annotations Start/stop, center/span, or CW frequency , source level, scale/div , reference level, marker data, softkey func- tions, warning and caution messages, trace identifica- tion, and pass/fail indication. Autoscale Automatically selects scale resolution and reference value to center the trace. Electrical delay Offset measured phase or group delay by a defined amount of electrical delay , in seconds. Frequency blanking Blank out all frequency information on the display . Requires an instrument preset to re-enable frequency information on the display . T itle Add custom titles (49 characters maximum) to the dis- play . Adjust display Customize the color and brightness of the data traces, memory traces, reference lines, graticules, text, and warning messages. Default colors can be recalled along with one set of user -defined display values. Storage Instrument state Up to 31 instrument states can be stored internally or recalled via the SA VE/RECALL menu. Instrument states include all control settings, active limit lines, active list frequency tables, memory trace data, active calibration coefficients, and custom display titles. Storage is in nonvolatile memory . T est sequences Six measurement sequences can be stored or recalled via the sequencing menu. Sequences may also be recalled from Preset menu. Sequence register 6 is part of nonvolatile storage and is not erased during a power cycle. If sequence 6 is titled AUTO, it will be executed when power is turned on. Disk drive Data, instrument states, user graphics, data plots and test sequences can be stored on internal floppy disk in MS-DOS ® or Hewlett-Packard's standard LIF formats. Data hardcopy Data plotting Hardcopy plots are automatically produced with HP- GL compatible digital plotters. The HP 8753ET/ES pro- vides Centronics, RS-232C, and GP-IB interfaces. Data listings Printouts of instrument data are directly produced with a printer such as the HP DeskJet or LaserJet. Select black & white or color print. For a list of compatible printers, consult our printer -compatibility guide W eb page at http://www .hp.com/go/pcg Configure plots Configure plots completely from the network analyzer by defining pen color and line type for data, text mark- ers, graticules, and memory traces. Functions Plot trace(s), graticule(s), marker(s), or text including operating and system parameters. Quadrants Plot entire display in one of four different quadrants of the plotter paper . System capabilities Limit lines Define test limit lines that appear on the display for go/no go testing. Lines may be any combination of hor- izontal, sloping lines, or discrete data points. Limit-test TTL output available for external control or indication. External source mode The receiver (input R) detects and phase-locks to any externally generated CW signal. Receiver inputs A and B will measure this same frequency for comparison or tracking measurements. Automatic The input signal frequency is counted and displayed. Manual Measures the input signal closest to the frequency specified by the user (within 0.5 to 5 MHz). T uned receiver T unes the receiver for a synthesized CW input signal at a precisely specified frequency . The time bases of the external RF source or sources must be tied to the external reference input (rear panel BNC). The built- in RF source is not used.
19 HP 8753ET/ES supplemental characteristics (continued) Frequency offset on/of f Sets the RF source to be swept at a fixed offset fre- quency above the receiver as required in a swept RF/IF , fixed LO, mixer test. Service menu Select the desired service test, service diagnostic, ser- vice or verification mode. T est sequences Description Create, edit, save or recall a series of front-panel key- strokes to automate a measurement. T est sequences may contain basic stimulus and measurement func- tions (frequency , power , parameter , format, scale) advanced operations (time domain, limit testing, dis- play marker values) and basic logical branching (IF limit test fails DO sequence 5 or GOSUB). Storage T est sequences can be stored internally to a disk drive and can be loaded from a computer over the GP-IB interface. Sequence 6 is saved in nonvolatile storage and can be used as an autostart routine when titled AUTO. General purpose input/output Read or write bits to the output port to control exter- nal devices such as part handlers. Eight output and five input TTL lines are available on the parallel port. Other functions P AUSE/continue, wait, title sequence, print sequence, duplicate sequence, pause and select. T ime-domain (Option 010) W ith the time-domain option, data from transmission or reflection measurements in the frequency domain are converted to the time domain using a Fourier transformation technique (chirp Z) and presented on the display . The time-domain response shows the mea- sured parameter value versus time. Markers may also be displayed in electrical length (or physical length if the relative propagation velocity is entered). T ime stimulus modes T wo types of time excitation stimulus waveforms can be simulated during the transformations, a step and an impulse. • Low-pass step This stimulus, similar to a traditional time-domain reflec- tometer (TDR) stimulus waveform, is used to measure low-pass devices. The frequency-domain data should extend from DC (extrapolated value) to a higher value. • Low-pass impulse This stimulus is also used to measure low-pass devices. • Bandpass impulse The bandpass impulse stimulates a pulsed RF signal (with an impulse envelope) and is used to measure the time-domain response of band-limited devices. W indows The windowing function can be used to modify (filter) the frequency-domain data and thereby reduce over- shoot and ringing in the time-domain response. Three types of windows are available: minimum, normal, and maximum. Gating The gating function can be used to selectively remove reflection or transmission time-domain responses. In converting back to the frequency-domain the effects of the responses outside the gate are removed. Remote programming Interface HP-IB interface operates to IEEE 488-1978 and IEC 625 standards and IEEE 728-1982 recommended prac- tices. Addressing The GP-IB address can be verified or set from the front panel via the local menu and can range from 0 to 30 decimal (factory set at 16). Pass control Allows the HP 8753ET/ES to request control of the GP-IB (when an active controller is present) whenever it needs to output to a plotter or printer . System controller Lets an HP 8753ET/ES become a controller on the GP-IB to directly control a plotter or a printer . T alker/listener Lets the HP 8753ET/ES become an GP-IB talker/listen- er when an external controller is present. T ransfer formats Binary (internal 48-bit floating-point complex format) ASCII 32- or 64-bit IEEE 754 floating-point format User -accessible graphics Using a subset of HP Graphics Language (HP-GL), vector or text graphics may be written on the HP 8753ET/ES via GP-IB. Up to 5 kbytes of data can be stored at one time (4 bytes per vector , 2 bytes per character). Interface function codes SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT0, C1, C2, C3, C10, E2
20 General characteristics Front panel connectors HP 8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal) HP 8753ES Option 011 test ports Connector type T ype-N female Impedance 50 ohms (nominal) HP 8753ES Option 075 test ports Connector type T ype-N female Impedance 75 ohms (nominal) HP 8753ET test ports Connector type T ype-N female Impedance 50 ohms (nominal) Probe power 15V ± 2% 400 mA (combined load for both probe connections) -12.6V ±5.5% 300 mA (combined load for both probe connections) Rear -panel connectors External reference frequency input (EXT REF INPUT) Frequency 1, 2, 5, and 10 MHz (±200 Hz at 10 MHz) Level -10 dBm to 20 dBm, (typical) Impedance 50 ohms Connector BNC (f) High-stability frequency reference output (Option 1D5) Frequency 10.0000 MHz Frequency stability ± 0.05 ppm (0 ˚C to 55 ˚C) Daily aging rate (after 30 days) <3x10 -9 /day Y early aging rate 0.5 ppm/year Output 0 dBm minimum Nominal output 50 Ω Connector BNC (f) impedance External auxiliary input (AUX INPUT) Input voltage limits -10V to 10V External AM input (EXT AM) ± 1 volt into a 5 k Ω resistor , 1 kHz maximum, resulting in approximately 8 dB/volt amplitude modulation. BNC (f) connector . External trigger (EXT TRIGGER) T riggers on a negative TTL transition or contact closure to ground. BNC (f) connector . T est sequence output (TEST SEQ) By default, this connector outputs a TTL end-of-sweep signal. It can also be programmed by the user in a test sequence to output a user -defined TTL signal. BNC (f) connector . Limit test output (LIMIT TEST) This connector outputs a TTL signal of the limit test results. Pass: TTL high. Fail: TTL low . BNC (f) connec- tor . T est-port bias input (BIAS CONNECT) (HP 8753ES only) Maximum voltage 30 VDC Maximum current (no degradation in RF specs) ± 200 mA Maximum current ± 1 A Connector BNC (f) VGA video output (EXT MON) This connector drives external VGA monitors. GP-IB This connector allows communications with compati- ble devices including external controllers, printers, plotters, disk drives, and power meters. Parallel port This 25-pin female connector is used with parallel peripherals. It can also be used as a general purpose I/O port. RS-232C This 9-pin male connector is used with serial peripherals. DIN keyboard This mini-DIN connector is used for adding an IBM PC-A T compatible keyboard. T est set interconnect This connector is used to connect an HP 8753ES Option 011 to the HP 85046A/B or 85047A test set. HP 8753ES analyzers without Option 011 can use sig- nal levels on this connector for sequencing or general purpose I/O applications. Internal memory T ypical data retention time with 3V , 1.2 Ah battery: At 25 ˚C 11904 days (32.6 years) At 40 ˚C 1244 days (3.4 years) At 70 ˚C 250 days (0.68 year) Line power 48 Hz to 66 Hz 115V nominal (90V to 132V) or 230V nominal (198V to 264V). 280 V A max. 1. F max is the upper frequency limit of the associated test set. 2. Degrees, specified as deviation from linear phase.
21 Environmental characteristics General conditions RFI and EMI susceptibility: defined by VDE 0730, CISPR Publication 11, and FCC Class B Standards. ESD (electrostatic discharge): must be eliminated by use of static-safe work procedures and an anti-static bench mat. The flexible rubber keypad protects key contacts from dust, but the environment should be as dust-free as possible for optimal reliability . Operating conditions T emperature (unless otherwise noted) 0˚ to 55 ˚C Humidity 5% to 95% at 40 ˚C (non-condensing) Altitude 0 to 4500 meters (15,000 feet) Non-operating storage conditions T emperature -40 ˚C to 70 ˚C Humidity 0 to 90% relative at 65 ˚C (non- condensing) Altitude 0 to 15,240 meters (50,000 feet) W eight (HP 8753ET/ES) Net 21 kg (46 lb) Shipping 35 kg (77 lb) Cabinet dimensions (HP 8753ET/ES) (These dimensions exclude front and rear panel protrusions.) 222 mm H x 425 mm W x 457 mm D (8.75 in x 16.75 in x 18.0 in) HP 8753ET/ES physical dimensions Measurement throughput summary The following table shows typical measurement times in milliseconds. T ypical time for completion (msec) Number of Points Measurement 51 201 401 1601 Start=1 GHz, Span=10 MHz, IFBW=6 kHz Uncorrected, 1-port calibration: 1 32 70 121 423 T wo-port calibration: 2 62 139 240 848 Start=30 kHz, Stop=3 GHz, IFBW=6 kHz Uncorrected, 1-port calibration 1 202 270 304 615 T wo-port calibration 2 402 540 607 1237 Start=30 kHz, Stop=6 GHz, IFBW=6 kHz Uncorrected, 1-port calibration 1 310 380 415 658 T wo-port calibration 2 618 757 829 1315 T ime domain (increase over uncorrected sweep time) 3 Conversion 12 42 86 378 Gating 14 40 80 349 GP-IB data transfer 4 Internal binary 10 16 21 58 ASCII 35 112 214 831 IEEE 754 floating point format 32-bit 11 19 28 83 64-bit 13 26 42 141 Options Harmonic measurements (Option 002) Measures amplifier 2nd and 3rd harmonics on a swept- frequency basis for fundamental signals above 16 MHz. Dynamic range (source at 10 dBm, receiver < 30 dBm) 40 dBc (minimum) Accuracy 5 ±1 dB (< 6 GHz) Step attenuator (Option 004) Provides source output power range from -85 to 10 dBm for HP 8753ET model. 6 GHz operation (Option 006) W ith Option 006, performance is extended to 6 GHz. When external source, tuned receiver or harmonic mode is used, the receiver is capable of measuring sig- nals up to 6 GHz. T ime domain (Option 010) T ransforms data from the frequency domain to the time domain using a Fourier transformation technique. High-stability frequency reference (Option 1D5) This option adds an ovenized 10-MHz frequency refer- ence output. It is connected to the external reference input on the rear panel. See the “General Characteristics” section for below specifications. 1. One-port calibration, with a 6 kHz IF bandwidth. Includes system retrace time, but does not include bandswitch time. T ime-domain gating is assumed off. 2. Same as footnote 1, but for an S21 measurement with full two-port calibration. Includes RF switching time. 3. Option 010 only , gating off. 4. Measured with an HP Omnibook 7100 266 MHz Pentium ® II computer . 5. Does not include error from the HP 8753T/ES source and receiver harmonics.
22 T est set specifications for Option 011 HP 85046A/B S-parameter test sets The HP 85046A/B S-parameter test sets provide the capability to measure reflection and transmission char- acteristics (including S-parameters) of two-port devices in either direction with a single connection. The test sets are controlled from the HP 8753ES Option 011 and include a programmable step attenua- tor . The frequency range of the HP 85046A 50-ohm test set is 300 kHz to 3 GHz. The HP 85046A has preci- sion 7-mm connectors. The frequency range of the HP 85046B 75-ohm test set is 300 kHz to 2 GHz. The HP 85046B has 75-ohm type-N(f) connectors. Both connectors can be adapted to other interfaces with the appropriate precision adapters. HP 85046A/B Specifications Note: Specifications that apply only to the HP 85046 are indicated in parentheses. Impedance 50 ohm (75 ohm) Frequency range 300 kHz to 3 GHz (300 kHz to 2 GHz) Directivity 35 dB to 1.3 GHz 30 dB to F max 1 T ypical tracking T ransmission magnitude, phase 2 0.3 MHz to 2.0 MHz ±1.5 dB, ±20˚ 2.0 MHz to F max ±1.5 dB, ±10˚ Reflection magnitude, phase 2 0.3 MHz to 2.0 MHz ±1.5 dB, ±25˚ 2.0 MHz to F max ±1.5 dB, ±10˚ Ef fective source match 0.3 MHz to 2.0 MHz 14 dB 2.0 MHz to 1.3 GHz 20 dB (17 dB) 1.3 GHz to F max 16 dB Nominal insertion loss Input to test port 14 dB 0.5 dB/GHz (19.5 dB 1 dB/GHz) Input to incident 18 dB 1.5 dB/GHz (18 dB 1.5 dB/GHz) Port 1, 2 to A, B 6.5 dB 1.0 dB/GHz (12 dB 0.5 dB GHz) T est set switch/repeatability ± 0.03 dB; ±0.01 dB 3 Max. operating level 20 dBm Damage level 30 dBm RF attenuator range 70 dB, 10 dB steps DC bias range ± 30 VDC, 200 mA 4 500 mA max DC bias connectors 50 ohm BNC (f) Includes four 190 mm (7.5 in) type-N cables and test set interconnect cable. Dimensions 90 mm H x 432 mm W x 553 mm D W eight 9.1 kg (20 lb) A standard HP 85046A/B test set contains a solid-state transfer switch, which allows continuous switching of power from port 1 to port 2 for full two-port error cor- rection. Option 009 replaces the transfer switch with a mechanical switch. This provides about 1.5 dB more power at the test port, but does not allow continuous switching, so the user must initiate updates of all four S-parameters for full two-port error correction. Also, the mechanical switch has relays that will wear out faster than the solid-state switch. Approximate lifetime of the mechanical switch is 1 million cycles. HP 85046A schematic 1. F max is the upper frequency limit of the associated test set. 2. Degrees, specified as deviation from linear phase. 3. T ypical performance. 4. Some degradation of RF specifications may occur .
23 HP 85047A S-parameter test set The HP 85047A S-parameter test set provides the capability to simultaneously measure the reflection and transmission characteristics of two-port devices in either direction with a single connection. This test set includes a frequency doubler that can be switched in by an HP 8753B/C Option 006 to measure 3 MHz to 6 GHz in a single sweep or switched out to measure 300 kHz to 3 GHz in a single sweep. The HP 8753ES Option 011 does not use the frequency doubler , so the full 300 kHz to 6 GHz range is available. This test set exhibits <5 dB insertion loss between the RF input and the test ports for as high as 15 dBm at the test port, and also includes a programmable step attenuator . There are two rear panel BNC outputs. One provides a TTL signal which indicates the result of a limit test. The second TTL output is controlled from the HP 8753ES test sequence function. Specifications Impedance 50 ohms Frequency range 300 kHz to 3 GHz and 3 GHz to 6 GHz (HP 8753B/C); 300 kHz to 6 GHz (HP 8753D/E/ES Opt. 006) Directivity 1 300 kHz to 1.3 GHz 35 dB 2 1.3 GHz to 3 GHz 30 dB 3 GHz to 6 GHz 25 dB T ypical tracking 1 T ransmission magnitude, phase 3 300 kHz to 3 GHz ±1.5 dB, ±10˚ 3 GHz to 6 GHz 0.5, -2.5 dB, ±20˚ Reflection magnitude, phase 3 300 kHz to 3 GHz ±1.5 dB, ±10˚ 3 GHz to 6 GHz ±1.5 dB, ±20˚ Source match 1 300 kHz to 1.3 GHz 20 dB 1.3 GHz to 3 GHz 16 dB 3 GHz to 6 GHz 14 dB Normal insertion loss Input to port 1,2 300 kHz to 3 GHz 4.0 dB 0.8 dB/GHz 3 GHz to 6 GHz 17.5 dB 0.8 dB/GHz Input to R 300 kHz to 3 GHz 19 dB 0.5 dB/GHz 3 GHz to 6 GHz 34 dB 0.5 dB/GHz Port 1,2 to A,B 16 dB T ypical Isolation 4 300 kHz to 3 GHz 100 dB 3 GHz to 6 GHz 90 dB T est port switch repeatability ± 0.03 dB; ±0.01 dB 4 Maximum operating level 20 dBm Damage level 30 dBm RF attenuator range 70 dB (10 dB steps) DC bias range ± 30 VDC, 200 mA, no degradation in RF specs, 1A max. RF connectors Port 1,2 7 mm precision All others 50 ohm type-N(f) Dimensions 90 mm H x 432 mm W x 553 mm D W eight 10 kg (22 lb) A standard HP 85047A test set contains a solid-state transfer switch, which allows continuous switching of power from port 1 to port 2 for full two-port error cor- rection. Option 009 replaces the transfer switch with a mechanical switch. This provides about 2.5 to 3.5 dB more power at the test port, but does not allow contin- uous switching, so the user must initiate updates of all four S-parameters for full two-port error correction. Also, the mechanical switch has relays that will wear out faster than the solid-state switch. Approximate lifetime of the mechanical switch is 1 million cycles. HP 85047A schematic 1. This can be greatly improved with accuracy enhancement. 2. Some degradation at environmental extremes below 600 kHz. 3. Degrees, specified as deviation from linear phase. 4. T ypical performance.
24 Accessories Calibration kits V ector accuracy enhancement procedures require that the systematic errors of the measurement system be characterized by measuring known devices (standards) on the system over the frequency range of interest. Return loss specifications or typical values are provid- ed where available for the terminations and adapters. HP 85031B 7-mm calibration kit Contains precision 7-mm standards used to calibrate the HP 8753ES for measurement of devices with preci- sion 7-mm connectors. Includes HP part number 7 mm short/open circuit 85031-60001 7 mm 50-ohm terminations (two each) 00909-60008 Specifications for terminations DC to 5 GHz RL ≥ 52 dB 5 to 6 GHz RL ≥ 46 dB HP 85032B 50-ohm type-N calibration kit Contains precision 50-ohm type-N standards used to calibrate the HP 8753ET/ES and 50-ohm test sets for measurement of devices with 50-ohm type-N connec- tors. Precision phase-matched 7-mm to type-N adapters are included for accurate measurements of non-insertable devices. Option 001 deletes the 7-mm to type-N adapters, which are not needed with the HP 8753ET . Includes HP part number N-male 50-ohm termination 00909-60009 N-female 50-ohm termination 00909-60010 N-male short circuit 85032-60008 N-female short circuit 85032-60009 N-female open circuit 85032-60012 N-male open circuit 85032-60007 7-mm to N-male adapter (two each) 85054-60009 7-mm to N-female adapter (two each) 85054-60001 Specifications for terminations: DC to 3 GHz RL ≥ 49 dB 2 to 3 GHz RL ≥ 46 dB 3 to 6 GHz RL ≥ 40 dB T ypical adapter characteristics: DC to 6 GHz RL ≥ 30 dB HP 85033D 3.5-mm calibration kit Contains a set of precision 3.5-mm standards to cali- brate the HP 8753ES and 50-ohm test sets for the measurement of devices with precision 3.5-mm and SMA connectors. Precision phase-matched 7-mm to 3.5-mm adapters are included for accurate measure- ments of non-insertable devices. Option 001 deletes the 7-mm to type-N adapters. Includes HP part number 3.5-mm-male 50-ohm termination 85033-60009 3.5-mm-female 50-ohm termination 85033-60010 3.5-mm-female short 85033-60014 3.5-mm-male short 85033-60013 3.5-mm-female open 85033-60012 3.5-mm-male open 85033-60011 7-mm to 3.5-mm female adapter (two) 1250-1747 7-mm to 3.5-mm male adapter (two) 1250-1746 Specifications for terminations: DC to 1.3 GHz RL ≥ 46 dB 1.3 to 3 GHz RL ≥ 44 dB 3 to 6 GHz RL ≥ 38 dB T ypical adapter characteristics: DC to 6 GHz RL ≥ 34 dB HP 85036B 75-ohm type-N calibration kit Contains a set of precision 75-ohm type-N standards to calibrate the HP 8753ES with a 75-ohm test set for measurement of devices with 75-ohm type-N connec- tors. Precision phased-matched adapters are included for accurate measurements of non-insertable devices. Includes HP part number N-male 75-ohm termination 00909-60019 N-female 75-ohm termination 00909-60020 N-female 75-ohm short 85036-60011 N-male 75-ohm short 85036-60012 N-female open 85032-20001 N-male open 85032-60007 N-male to N-male 75-ohm adapter 85036-60013 N-female to N-female 75-ohm adapter 85036-60014 N-male to N-female 75-ohm adapter 85036-60015 Specifications for terminations: DC to 2 GHz RL ≥ 46 dB 2 to 3 GHz RL ≥ 40 dB
25 HP 85038 family of calibration kits The HP 85038A calibration kit contains open circuits, short circuits, and terminations with both male and female 7-16 connectors. More economical versions of this kit are available: the HP 85038M contains only male devices, and the HP 85038F provides only female devices (see following table). Each kit contains a flop- py disk with the calibration kit definition for use with HP 8712, 8714, 8753, 8719, 8720, 8722, and 8510C net- work analyzers. Contents of 7-16 Calibration Kits Description HP Part Number HP 85038A HP 85038M HP 85038F Open circuit, female 85038-80002 X X Open circuit, male 85038-80003 X X Short circuit, female 85038-80004 X X Short circuit, male 85038-80005 X X T ermination, female 85038-80006 X X T ermination, male 85038-80007 X X 7-16 adapter , female-to-female 11906-80016 X 7-16 adapter , male-to-male 11906-80015 X Calibration kit data disk 85038-10001 X X X T orque wrench 8710-2175 X Open-end wrench 8710-2174 X Specifications Frequency coverage dc to 7.5 GHz (usable to 8 GHz) Nominal impedance 50 ohms Open circuits Reflection phase ± 1 degree from nominal Short circuits Reflection phase ±1 degree from nominal T erminations: Return loss > 40 dB 7-16 to 7-16 adapters Return loss > 40 dB (VSWR < 1.02) 7-16 to 50 ohm type-N Return loss > 36 dB adapters (VSWR < 1.03) 7-16 to 7-mm adapters Return loss > 38 dB (VSWR < 1.025) 7-16 to 3.5-mm adapters Return loss > 34 dB (VSWR < 1.04) HP 11906 family of adapter kits The HP 11906 family consists of four adapter kits: HP 11906A 7-16 to 7-16 adapter kit Description Quantity HP Part Number 7-16 male-to-male adapter 1 11906-80015 7-16 female-to-female adapter 1 11906-80016 7-16 male-to-female adapter 2 11906-80017 HP 11906B 7-16 to 50-Ohm type-N adapter kit Description Quantity HP Part Number T ype-N male to 7-16 male adapter 1 11906-80007 T ype-N female to 7-16 female adapter 1 11906-80008 T ype-N female to 7-16 male adapter 1 11906-80009 T ype-N male to 7-16 female adapter 1 11906-80010 HP 11906C 7-16 to 7-mm adapter kit Description Quantity HP Part Number 7-mm to 7-16 male adapter 2 11906-80012 7-mm to 7-16 female adapter 2 11906-80013 HP 11906D 7-16 to 3.5-mm adapter kit Description Quantity HP Part Number 3.5-mm male to 7-16 male adapter 1 11906-80002 3.5-mm male to 7-16 female adapter 1 11906-80005 3.5-mm female to 7-16 male adapter 1 11906-80004 3.5-mm female to 7-16 female adapter 1 11906-80003 HP 85039B type-F calibration kit Contains a set of 75-ohm type-F standards to calibrate the HP 8753ES with a 75-ohm test set for the mea- surement of devices with type-F connectors. Includes HP part number F-male 75-ohm termination 85039-60007 F-female 75-ohm termination 85039-60004 Specifications for termination DC to 1 GHz RL ≥ 45 dB 1 to 3 GHz RL ≥ 38 dB F-male 75-ohm short 85039-60008 F-female 75-ohm short 85039-60003 F-male 75-ohm open 85039-60009 F-female 75-ohm open 85039-60005 F-female to F-female 75-ohm adapter 85039-60002 F-male to F-male 75-ohm adapter 85039-60006 T ypical type-F adapter characteristics DC to 1 GHz RL ≥ 40 dB 1 to 3 GHz RL ≥ 32 dB F-female to N-male 75-ohm adapter 85039-60013 F-male to N-female 75-ohm adapter 85039-60011 T ypical type-F to type-N adapter characteristics DC to 1 GHz RL ≥ 38 dB 1 to 3 GHz RL ≥ 32 dB
26 V erification kits Measuring known devices other than the standards used in calibration is an easy way to verify the proper operation of an HP 8753ES measurement system. HP offers verification kits which include devices, with data, for verifying the error -corrected measurements of an HP 8753ES and 50-ohm test sets. HP 85029B 7-mm verification kit Contains a set of precision 7-mm devices, with data traceable to NIST* for verifying the calibrated perfor- mance of an HP 8753E measurement system. The HP 85031B 7-mm calibration kit is required for complete verification. T est-port return cables Hewlett-Packard offers high-quality RF cables to con- nect the HP 8753ET/ES and test sets to devices under test. These cables offer excellent RF shielding for high- dynamic-range measurements. HP 11851B 50-ohm type-N RF cable kit Recommended for use with the HP 11850C/D three- way power splitters. Kit includes three phase-matched 610-mm (24-in.) cables and one 860-mm (34-in.) cable. Return loss > 24 dB to 3 GHz Phase tracking ±4˚ at 1.3 GHz HP 8120-5639 50-ohm type-N RF cable A 50-ohm type-N test port cable (both connectors are type-N male). Return loss > 30dB to 6 GHz HP 11857B 75-ohm type-N test port return cables A pair of 610-mm (24-in.) test port return cables for use with the HP 8753ES or HP 85046B 75-ohm S-para- meter test set. Return loss > 24 dB to 2 GHz Phase tracking ±2˚ at 1.3 GHz HP 11857D 7-mm test-port return cables A pair of 610-mm (24-in.) test port return cables for use with the HP 8753ES with the HP 85046A and 85047A S-parameter test sets. These cables can be used with connector types other than 7-mm with the appropriate precision adapters. Return loss > 24 dB to 3 GHz > 20 dB to 6 GHz Phase tracking ±2˚ at 1.3 GHz HP 11857F type-F test-port return cables T ype-F test port cable set (75-ohm). Includes one cable with type-N(male) to type-F(male) connectors and one cable with type-N(male) to type-F(female) connectors. HP 11850C/D three-way power splitters HP 11850C HP 11850D Impedance 50 ohms 75 ohms Frequency range DC to 3 GHz DC to 2 GHz T racking ±25 dB, ±3˚ ±2 dB, ±2.5˚ Equivalent source match 30 dB at 1.3 GHz 30 dB at 1.3 GHz (ratio or leveling) 20 dB at 3 GHz 20 dB at 2 GHz Nominal 9.5 dB 1 dB/GHz 7.8 dB insertion loss Input port match DC to 1.3 GHz 20 dB 20 dB 1.3 GHz to F max 10 dB 10 dB Maximum operating level 20 dB 20 dB Damage level 30 dB 30 dB RF connectors RF input 50-ohm type-N (f) 50-ohm type-N(f) All others 50-ohm type-N (f) 75-ohm type-N(f) Includes Three HP 11852B 50-to 75-ohm minimum-loss pads Recommended accessories HP 11851B RF cable kit HP 11667A 50-ohm power splitter Frequency range DC to 18 GHz T ypical insertion loss 6 dB Equivalent source match DC to 4 GHz 26 dB 4 GHz to 8 GHz 21 dB 8 GHz to 18 GHz 17 dB T racking (between output arms) DC to 4 GHz ±15 dB 4 GHz to 8 GHz ±2 dB 8 GHz to 18 GHz ±25 dB Maximum operating level ± 27 dBm Connectors 50-ohm type-N (f) Option 001 RF input type-N (m) Outputs type-N (f) Option 002 RF input type-N (f) on RF Outputs precision 7-mm Dimensions 46 mm H x 52 mmW x 19 mm D (1.8 x 2.0 x 0.7 in) Recommended accessories HP 11851B RF cable kit HP 11852B 50-ohm to 75-ohm minimum loss pad Frequency range DC to 3.0 GHz Nominal insertion loss 5.7 dB Return loss 300 kHz to 2 GHz 32 dB 2 GHz to 3 GHz 27 dB Maximum input power 250 mW ( 24 dBm)
27 HP 11852B 50-ohm to 75-ohm minimum loss pad (continued) Connectors 50-ohm type-N (f) to 75-ohm type-N (m) standard Option 004, 50-ohm type-N (m) to 75-ohm type-N (f) Dimensions 14-mm D x 70-mm L (0.56 in x 2.75 in) W eight Net 0.1 kg (0.316 lb) 50-ohm accessory kits The HP 11853A 50-ohm type-N and the HP 11854A 50-ohm BNC accessory kits provide the RF compo- nents generally required when using either the HP 85046A, 85047A or 11850C with the HP 8753ES Option 011 when measuring devices having 50-ohm type-N or BNC connectors. These kits are supplied with a storage case. HP 11853A 50-ohm type-N accessory kit Includes HP part number T ype-N (f) short HP 11511A T ype-N (m) short HP 11512A T ype-N (m) to N (m) adapter 1250-1475 T ype-N (f) to N (f) adapter 1250-1472 HP 11854A 50-ohm BNC accessory kit Includes HP part number T ype-N (m) to BNC female adapter 1250-1476 T ype-N (m) to BNC male adapter 1250-1473 T ype-N (f) to BNC male adapter 1250-1477 T ype-N (f) to BNC female adapter 1250-1474 BNC (m) short 1250-0929 75-ohm accessory kits The HP 11855A 75-ohm type-N and the HP 11856A 75-ohm BNC accessory kits provide the RF compo- nents generally required when using either the HP 85046B or the HP 11850D power splitter with the HP 8753ES Option 011 when measuring devices having 75-ohm type-N or BNC connectors. These kits are sup- plied with a storage case. HP 11855A 75-ohm type-N accessory kit Includes HP part number T ype-N (f) short 1250-1531 T ype-N (m) short 1250-1530 T ype-N (m) to N (m) adapter 1250-1528 T ype-N (f) to N (f) adapter 1250-1529 T ype-N (m) termination 1250-1532 HP 11856A 75-ohm BNC accessory kit Includes HP part number T ype-N (m) to BNC (f) adapter 1250-1535 T ype-N (m) to BNC (m) adapter 1250-1533 T ype-N (f) to BNC (m) adapter 1250-1534 T ype-N (f) to BNC (f) adapter 1250-1536 BNC (m) short 1250-0929 BNC (m) termination 11652-60010 RF limiter Externally attaches to one or both ports of the analyz- er . Provides protection against potential high-power transients from external devices. Specifications HP 11930A 7-mm RF limiter Frequency range DC to 6 GHz Nominal insertion loss DC to 3 GHz 1.0 dB 3 GHz to 6 GHz 1.5 dB Return loss DC to 3 GHz 22 dB 3 GHz to 6 GHz 20 dB Maximum input power 3W Maximum DC 30 V , 350 mA HP 11930B 50-ohm type-N RF limiter 2 Frequency range 5 MHz to 6 GHz Nominal insertion loss 5 MHz to 3 GHz 1.0 dB 1 3 GHz to 6 GHz 1.5 dB Return loss 5 MHz to 3 GHz 21 dB 1 3 GHz to 6 GHz 17 dB Maximum input power 3W HP 85024A high-frequency probe T wo probes may be powered directly from the front panel of the HP 8753ET/ES. Refer to technical specifications, HP literature #5954-8393, for more information. Specifications Input capacitance (at 500 MHz) <0.7 pF (nominal) Input resistance 1 Megohm (nominal) Bandwidth 300 kHz to 3 GHz Gain (at 500 MHz) 0 dB ±1 dB Frequency response 300 kHz to 1 GHz ±1 dB 1 GHz to 3 GHz 2, -3 dB Input voltage for < 1 dB compression 0.3 V Supplement characteristics Noise figure <100 MHz < 50 dB 100 MHz to 3 GHz < 25 dB Includes HP part/model number T ype-N (m) adapter HP 11880A 10:1 divider HP 11881A Spare 12 mil probes 85024-20012 2.5-inch ground lead 01223-61302 Hook tip HP 10229 Spanner tip 5060-0549 Probe tip nut driver 8710-1806 1. Return loss and insertion loss limited below 16 MHz by series capacitor . 2. Internal bias tees cannot be used with this limiter .
28 HP 8347A RF amplifier The HP 8347A RF amplifier delivers increased power across a 300 kHz to 3 GHz frequency range. The HP 8347A provides leveled output power without using an external coupler and detector , since these parts are built-in. The external ALC can be directly connected to the external AM input on the HP 8753ET/ES. This capability is especially useful for achieving high dynamic range measurements at faster sweep rates. Specifications Frequency 100 kHz to 3 GHz Gain 25 dB minimum Output power (leveled) 5 dBm to 20 dBm (adjustable) Maximum output power 24 dBm (typical) Leveled power flatness ± 1.5 dB Impedance 50 ohms nominal SWR Input 2.2:1 max Output 1.6:1 (ALC on) Spectral purity Harmonics -20 dBc at dBm Third-order intercept 30 dBm (nominal) T ypical noise figure 100 MHz to 3 GHz 13.5 dB RF connectors T ype-N female Dimensions 102-mm H x 213-mm W x 297-mm D (4.0 in x 8.4 in x 11.7 in) W eight net 3.5 kg (7.7 lb) MS-DOS is a U.S. registered trademark of Microsoft Corporation. Pentium is a U.S. registered trademark of Microsoft Corporation. For more information about Hewlett-Packard test and measure- ment products, applications, services, and for a current sales office listing, visit our web site, http://www .hp.com/go/tmdir . Y ou can also contact one of the following centers and ask for a test and measurement sales representative. United States: Hewlett-Packard Company T est and Measurement Call Center P .O. Box 4026 Englewood, CO 80155-4026 1 800 452 4844 Canada: Hewlett-Packard Canada Ltd. 5150 Spectrum W ay Mississauga, Ontario L4W 5G1 (905) 206 4725 Europe: Hewlett-Packard European Marketing Centre P .O. Box 999 1180 AZ Amstelveen The Netherlands (31 20) 547 9900 Japan: Hewlett-Packard Japan Ltd. Measurement Assistance Center 9-1, T akakura-Cho, Hachioji-Shi, T okyo 192, Japan T el: (81) 426-56-7832 Fax: (81) 426-56-7840 Latin America: Hewlett-Packard Latin American Region Headquarters 5200 Blue Lagoon Drive, 9th Floor Miami, Florida 33126, U.S.A. (305) 267 4245/4220 Australia/New Zealand: Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130, Australia 1 800 629 485 Asia Pacific: Hewlett-Packard Asia Pacific Ltd 19/F Cityplaza One 1111 King's Road T aikoo Shing, Hong Kong tel: 852-2599-7777 fax: 852-2506-9285 Data Subject to Change Copyright © 1999 Hewlett-Packard Company Printed in U.S.A. 8/99 5968-5160E